| Model Number |
1012453-08 |
| Medical Device Problem Codes |
Leak/Splash (1354); Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported complaints appear to be related to operational context.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a de novo lesion located in the mid-right coronary artery that was moderately tortuous and 90% stenosed.The 4.0x8mm nc trek balloon dilatation catheter was advanced to the lesion without issue; however after the first inflation, a leak of the contrast agent was noted from the balloon.After removal, a pinhole rupture was noted.There was no adverse patient effect or a clinically significant delay in the procedure.A 4.0x12mm nc trek balloon dilatation catheter was used to complete the procedure.No additional information was provided.
|
| |
|