Model Number 3544251
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 12/01/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a 30-year-old caucasian female patient underwent a primary breast augmentation surgery with implantation of a 425cc mentor memorygel breast implant prosthesis.Post-operatively, the patient was diagnosed with a ruptured right breast implant during an in-office visit as the medical professional noted a loss of fullness.As a result, the patient underwent removal and replacement with 485cc natrelle inspira cohesive breast implants on (b)(6), 2022.
 
Additional Manufacturer Narrative
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.Reason for device explant and/or reoperation: rupture.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On september 23, 2022, mentor completed an evaluation on the returned device.Mentor conducted a visual inspection and microscopic examination of the device.Visual analysis of the returned sample revealed that the breast implant was found to be ruptured and received in two parts.A microscopic examination was performed, and the cause of the tear could not be identified.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: rupture can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.The following things may cause the implant to rupture: damage by surgical instruments; stressing the implant during implantation and weakening it; folding or wrinkling of the implant shell; excessive force to the chest (e.G.During closed capsulotomy); trauma; compression during mammographic imaging; and severe capsular contracture.Breast implants may also simply wear out over time.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
michelleann garcia
3041 skyway circle north
irving, TX 75038
646591-7981
MDR Report Key15316630
Report Number1645337-2022-10117
Device Sequence Number14829031
Product Code FTR
UDI-Device Identifier00081317003885
UDI-Public00081317003885
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2019
Device Model Number3544251
Device Catalogue Number3544251
Device Lot Number6828816
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/26/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/10/2022
Supplement Date Received by Manufacturer09/23/2022
Initial Report FDA Received Date08/29/2022
Supplement Report FDA Received Date09/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/28/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age30 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-