| Model Number |
ORT300 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/30/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The suspected source of the fluid leak is a valve component within the hydraulic manifold.No hydraulic pressure loss or impact on table function has been observed.A followup report will be submitted once the investigation is complete.
|
| |
|
Event or Problem Description
|
|
During routine preventive maintenance, an imris customer service engineer observed several small drops of hydraulic fluid on outside of the hydraulic lines of the ort300 operating room table.No issues were reported or observed with respect to hydraulics controlled table functions.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
An imris customer service engineer made multiple site visits to address the preventive maintenance finding of a small hydraulic fluid leak from the operating room table.The source of the leak was not definitively determined, at a followup visit no fluid outside the hydraulic system was observed.Multiple suspected hydraulic system components were replaced as a preventive measure, including a flow control valve, a hose, and multiple fittings and unions.Hydraulic system pressure and fluid levels were measured to be within specification, and weighted drift testing was performed with passing results.No impact to table function was reported or observed.
|
| |
|