Model Number 6720-0435
Medical Device Problem Codes Unstable (1667); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 08/08/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Right hip revision for instability.Stem, insert, and head were revised.Per response: removal of stem was not planned during revision it was removed however as a result of intra-op event, "when the constrained inserts failed, stem removal was necessary".
 
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
Right hip revision for instability.Stem, insert, and head were revised.Per response: removal of stem was not planned during revision it was removed however as a result of intra-op event, "when the constrained inserts failed, stem removal was necessary.".
 
Additional Manufacturer Narrative
Reported event: an event regarding revision involving an accolade stem that was mated with a lfit v40 cocr head was reported.The event was not confirmed.Method & results: product evaluation and results: visual inspection: the device was not returned but photographs were provided for review.A review of the provided photographs indicated the device is covered in blood.Material analysis, functional, and dimensional inspections could not be performed as the device was not returned.Clinician review: a review of medical records with a clinical consultant indicated "revision of a tha can be confirmed by the intra-operative photographs.There is no documentation regarding the cause for revision being instability.Nor was any documentation provided indicating evidence of intra-operative failure as noted in the pi.The root cause for the revision surgery cannot be determined from this limited documentation.Should additional information become available i would be happy to further this assessment." product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the reported accolade stem was mated with an lfit v40 cocr head.The cocr head has been identified to be within scope of nc and capa.No further investigation for this event is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIZE 4 ACCOLADE II 132 DEG
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
diana rogers
210 centennial avenue
centennial park, elstree
borehamwood, NJ WD6 3-SJ
UK   WD6 3SJ
2018315000
MDR Report Key15317039
Report Number0002249697-2022-01264
Device Sequence Number5206209
Product Code LPH
UDI-Device Identifier04546540664471
UDI-Public04546540664471
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K143085
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2018
Device Model Number6720-0435
Device Catalogue Number6720-0435
Device Lot Number43367704
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/08/2022
Supplement Date Received by Manufacturer11/02/2022
Initial Report FDA Received Date08/29/2022
Supplement Report FDA Received Date11/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/24/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age75 YR
Patient SexFemale
-
-