The following devices were also listed in this report: modular dual mobility insert; cat# 626-00-42e; lot# 86241502; restoration adm x3 ins 28/48; cat# 1236-2-848 ; lot# 516378; delta v-40 ceramic head 28/+4; cat# 6570-0-228; lot# 83640702; tridentii tritanium cluster54e; cat# 702-04-54e; lot# 84349001a.It cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.Reported event: an event regarding infection involving a accolade stem was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate the devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot or sterile lot referenced.Conclusions: it was reported that the patient was revised due to infection.All stryker products sold as sterile are validated to a minimum sterility assurance level sal of 10^-6 in accordance with applicable iso standards.The exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.Device not returned.
|