AMO PUERTO RICO MFG. INC. TECNIS IOL; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number DIU525
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Blurred Vision (2137); Visual Impairment (2138)
Date of Event 05/31/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a post-operative (op) patient was having anisometropia and diplopia due to residual astigmatism.The intraocular lens (iol) was eventually explanted from the patient¿s right eye.Another lens, model diu150, 18.0 diopter was used as a replacement.Sutures on the main wound were used and no other interventions indicated.There was no injury and the patient¿s outcome were reported to be good.Allergic reaction to post-op medication reported, but they switched medications and the patient is doing much better.No further information was provided.
 
Additional Manufacturer Narrative
Device evaluation: the device was not returned at the manufacturing site; therefore, product testing could not be performed, and the customer¿s reported complaint could not be verified.Manufacturing record review: the manufacturing records review was performed, and no nonconformity report was found as part of this manufacturing records review.The product was manufactured and released according to specification.The search revealed that no other complaints for this production order number have been received.Conclusion: as a result of the investigation there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Additional information: section d9: device available for evaluation? yes.Section d9: date returned to manufacturer: sep 20, 2022.Section h3: device evaluated by manufacturer: yes.Device evaluation: the complaint lens was received in a non-johnson & johnson sterilization pouch.Visual inspection under magnification revealed that the lens was received cut in half and with a detached haptic.No defects that could contribute to visual issues were observed.Based on the return condition of the lens no further evaluation could be performed.The complaint issue ¿blurry vision¿ and ¿diplopia-persistent¿ was not confirmed.The other observed issues during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Therefore, there is no indication of a product deficiency or product malfunction.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key15317259
Report Number3012236936-2022-02292
Device Sequence Number5230219
Product Code MJP
UDI-Device Identifier05050474749375
UDI-Public(01)05050474749375(17)250222
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/22/2025
Device Model NumberDIU525
Device Catalogue NumberDIU525U140
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/01/2022
Supplement Date Received by Manufacturer09/29/2022
Initial Report FDA Received Date08/29/2022
Supplement Report FDA Received Date10/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient Weight54 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-