MEDTRONIC HEART VALVES DIVISION CONTEGRA; PULMONIC VALVED CONDUIT Back to Search Results
Model Number 200H14
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Obstruction/Occlusion (2422)
Date of Event 07/25/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Medtronic received information that 1 year and 1 month post implant of this 14mm pulmonary valved conduit in a (b)(6) pediatric patient, it was explanted and replaced with a 14mm conduit of the same model.The reason for replacement was reported to be dilation of the conduit due to a distal obstruction.It was mentioned that there was dense fibrous reaction at the pulmonary artery confluence.It was stated that this was fully dissected.It was noted that a transesophageal echocardiogram performed after replacement of the conduit showed a gradient of approximately 20 to 25mmhg in the proximal right pulmonary artery.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
No product was returned.Without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTEGRA
Common Device Name
PULMONIC VALVED CONDUIT
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer (Section G)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key15317578
Report Number2025587-2022-02404
Device Sequence Number12234260
Product Code MWH
UDI-Device Identifier00643169616905
UDI-Public00643169616905
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
H020003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/12/2023
Device Model Number200H14
Device Catalogue Number200H14
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/04/2022
Initial Report FDA Received Date08/30/2022
Date Device Manufactured11/12/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age13 MO
Patient SexFemale
-
-