| Model Number |
301803MS |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).This report is being submitted pursuant to the provisions of 21 cfr, part 4.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc.Or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that an unknown patient underwent an atrial fibrillation (afib) ablation procedure with a preface® guiding sheath with multipurpose curve.There was a valve physical abnormality.There was a hemostasis valve failure, valve physical abnormality.Reverse blood was present at the time of the septal puncture into the left atrium.This occurred 1 hour and 30 minutes after insertion of the sheath, at the time of the septal puncture in the left atrium.The catheter was exchanged to resolve the event, and the procedure was continued.The sheath was not being used on the patient.The complaint product was replaced to another sheath.Blood return was observed.No patient consequences were reported.Hemostatic valve separation is mdr-reportable.
|
| |
|
Additional Manufacturer Narrative
|
|
On 5-sep-2022, bwi received information indicating that the sheath was inserted into the patient when the complaint issue happened.On 8-sep-2022, the bwi product analysis lab received the device for evaluation.The analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On 5-oct-2022, the product investigation was completed.It was reported that an unknown patient underwent an atrial fibrillation (afib) ablation procedure with a preface® guiding sheath with multipurpose curve.There was a valve physical abnormality.Device evaluation details: visual analysis of the returned sample revealed no anomalies in the components returned.The functional and microscopic analyses were found within specifications.A manufacturing record evaluation was performed for finished device number 18079353, and no internal action related to the complaint was found during the review.The complaint reported by the customer as ¿hemostasis valve/hub ¿ leakage - during use¿ was not confirmed since during functional test no leaks were noted on the valve/hub.The exact cause of the reported event could not be conclusively determined during the analysis.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|
| |
|