Model Number M0035421020
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/02/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a coil embolization procedure was performed to treat cerebral aneurysm in a-com.While advancing the subject coil, physician felt resistance in the middle of the microcatheter.He was unable to advance or retrieve the subject coil and while doing so, the subject coil got prematurely detached.The subject coil was removed safely from the patients vascular anatomy.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Additional Manufacturer Narrative
The device is not available to the manufacturer.
 
Additional Manufacturer Narrative
Due to the automated mes system there are controls in the manufacturing process to ensure the product met specifications upon release.Visual/microscopic inspection: the main coil was seen to be detached.The main coil was seen to be kinked/bent and stretched.The coil delivery wire was seen to be kinked/bent.The coil introducer sheath was intact.Functional inspection: due to the detachment of the main coil, a functional inspection was not possible.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported event 'main coil prematurely detached/separated during use' was confirmed during analysis.The reported event 'coil in catheter friction' and 'coil introducer sheath friction' could not be confirmed; however, the analysis results are consistent with the reported event.The device failed to meet specifications when received for complaint investigation based on the analyzed anomalies noted to the device.The device was analyzed.The coil delivery wire was found to be kinked/bent.The main coil was found to be kinked/bent and stretched, and main coil was found to be prematurely detached/separated during use.An assignable cause of procedural factors will be assigned to the as reported 'coil introducer sheath friction, coil in catheter friction, as reported/ as analyzed 'main coil prematurely detached/separated during use, and as analyzed 'main coil kinked/bent' and 'main coil stretched', as the defects appears to be associated with a product that met stryker design and manufacturing specifications and was used in accordance with the dfu, but performance was limited due to procedural factors during use.An assignable cause of 'handling damage' will be assigned to the as analyzed 'coil delivery wire kinked/bent' as the defect appears to be associated with handling of the product or portion of the product upon removal of the product from the packaging.
 
Event or Problem Description
It was reported that a coil embolization procedure was performed to treat cerebral aneurysm in a-com.While advancing the subject coil, physician felt resistance in the middle of the microcatheter.He was unable to advance or retrieve the subject coil and while doing so, the subject coil got prematurely detached.The subject coil was removed safely from the patients vascular anatomy.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET 360 NANO 1MM X 2CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key15318385
Report Number3008881809-2022-00418
Device Sequence Number5505076
Product Code HCG
UDI-Device Identifier04546540697967
UDI-Public04546540697967
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/07/2025
Device Model NumberM0035421020
Device Catalogue NumberM0035421020
Device Lot Number23518578
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/02/2022
Supplement Date Received by Manufacturer10/19/2022
Initial Report FDA Received Date08/30/2022
Supplement Report FDA Received Date10/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/09/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MICROCATHETER (UNKNOWN)
Patient SexUnknown
-
-