• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number SONATA
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 12/12/2021
Type of Reportable Event Malfunction
Event or Problem Description
In december of 2021 poor auditory response with the device was observed.The user was re-implanted.
 
Additional Manufacturer Narrative
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Event or Problem Description
In (b)(6) 2021 poor auditory response with the device was observed.The user was re-implanted.
 
Additional Manufacturer Narrative
Conclusion: overload fractures in the active electrode which are consistent with an external mechanical impact were determined to be the root cause of device failure.The problems given in the recipient report appear to match the damage found.Other damages found during device investigation are most likely related to the explantation surgery.This is a final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM
Common Device Name
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
57788
MDR Report Key15318495
Report Number9710014-2022-00628
Device Sequence Number17450750
Product Code MCM
UDI-Device Identifier09008737049157
UDI-Public(01)09008737049157
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/01/2022
Device Model NumberSONATA
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/29/2022
Supplement Date Received by Manufacturer08/29/2022
Initial Report FDA Received Date08/30/2022
Supplement Report FDA Received Date11/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/01/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age35 MO
Patient SexFemale
-
-