• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA QUADRA-H CEMENTLESS, HA COATED LAT STEM SIZE 1, SHORT NECK; HIP CEMENTLESS STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA QUADRA-H CEMENTLESS, HA COATED LAT STEM SIZE 1, SHORT NECK; HIP CEMENTLESS STEM Back to Search Results
Model Number 01.12.31SN
Medical Device Problem Codes Device Appears to Trigger Rejection (1524); Osseointegration Problem (3003)
Health Effect - Clinical Codes Bone Fracture(s) (1870); Inadequate Osseointegration (2646)
Date of Event 08/08/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 08-aug-2022: lot 182707: 38 items manufactured and released on 20-jul-2018.Expiration date: 2023-07-11.No anomalies found related to the problem.To date, 29 items of the same lot have been sold with no similar reported case during the period of review.
 
Event or Problem Description
At about 2 months after the primary surgery, revision was performed due to implant loosening probably caused by a suspected fracture in the femur bone.Any fracture was noticed during the primary surgery.During the revision surgery a fracture was noticed but the reason is unknown.The surgeon revised successfully the head with a ceramic head and stem with a quadra-c stem and implanted a competitor plate.
 
Additional Manufacturer Narrative
Corrected data d4: the correct device reference is (b)(6).G3: date received by manufacturer is 08-05-2022., h6: medical device proble code is 3003.Additional info d4: the device expires on 07-11-2023.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA-H CEMENTLESS, HA COATED LAT STEM SIZE 1, SHORT NECK
Common Device Name
HIP CEMENTLESS STEM
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ   6874
Manufacturer Contact
stefano baj
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key15319223
Report Number3005180920-2022-00668
Device Sequence Number5203708
Product Code JDI
UDI-Device Identifier07630030802713
UDI-Public07630030802713
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K082792
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number01.12.31SN
Device Catalogue Number01.12.31SN
Device Lot Number182707
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/05/2022
Supplement Date Received by Manufacturer08/05/2022
Initial Report FDA Received Date08/30/2022
Supplement Report FDA Received Date08/30/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/20/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
Patient SexFemale
Patient Weight84 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-