• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEGADYNE MEDICAL PRODUCTS, INC. 10FT SMOKE PENCIL UCONN; RETURN ELECTRODE CABLE, DUAL PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEGADYNE MEDICAL PRODUCTS, INC. 10FT SMOKE PENCIL UCONN; RETURN ELECTRODE CABLE, DUAL PLATE Back to Search Results
Model Number 251010J
Medical Device Problem Code Flaked (1246)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/12/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an ent procedure that the device was plugged into the generator the surgeon activated the device and an inch and a half from the patient and a ¿ball of flame¿ appeared at the tip of the device.The entire set up including the generator, grounding pad, handpiece and pencil were replaced to continue with the procedure.The procedure was completed with no further issues.There were no adverse consequences for the patient.
 
Additional Manufacturer Narrative
(b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.No lot or batch number was provided therefore a device history could not be done.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 9/29/2022.Investigation summar: the product was returned to for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the 251010j device was received with no apparent damage.The instrument was tested with a generator and no alarms was observed during the analysis.There were no anomalies noted with the functionality of the device.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reached on the cause of the reported event.The event described could not be confirmed as the device performed without any difficulties noted.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 9/15/2022 additional information was requested, and the following was obtained: 1.What generator and return electrode were being used? 2.What were the power settings on the generator? unsure of power settings on generator 3.What mode was being used (cut or coag)? unsure of if it was cut or coag 4.Did it arc to something metal in the surgical field? no arcing 5.Any flammable gases present? no flammable gas.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
10FT SMOKE PENCIL UCONN
Common Device Name
RETURN ELECTRODE CABLE, DUAL PLATE
Manufacturer (Section D)
MEGADYNE MEDICAL PRODUCTS, INC.
4545 creek road
cincinnati OH 45242
Manufacturer (Section G)
MEGADYNE MEDICAL PRODUCTS, INC.
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
329348013
MDR Report Key15319377
Report Number1721194-2022-00076
Device Sequence Number12142094
Product Code GEI
UDI-Device Identifier10614559105573
UDI-Public10614559105573
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K982130
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number251010J
Device Catalogue Number251010J
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/17/2022
Supplement Date Received by Manufacturer08/30/2022
09/21/2022
Initial Report FDA Received Date08/30/2022
Supplement Report FDA Received Date09/15/2022
09/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-