MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL ENTERPRISE2 4MMX30MM; INTRACRANIAL NEUROVASCULAR STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL ENTERPRISE2 4MMX30MM; INTRACRANIAL NEUROVASCULAR STENT Back to Search Results
Model Number ENF403012
Medical Device Problem Code Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/12/2022
Type of Reportable Event Malfunction
Event or Problem Description
As reported by the field, during a stent assist coil embolization, an eu 4.5x28mm stent 12 mm dw tip (enc452812, 6920544) arrived at the target position and the physician tried to release it, but the stent could not be opened.The doctor changed to a new 4mmx30mm enterprise 2 intracranial neurovascular stent (encr403012, 6610181) but the same issue occurred.The third stent was switched to complete the surgery.There was no patient injury reported.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Information regarding patient weight, height, medical history, race, and ethnicity was not reported.Initial reporter phone: (b)(6).The device is available to be returned for evaluation and testing.However, it has not been received to date as indicated as ¿other¿ in this section as the reason for non-evaluation.If the device returns, a device investigation will be performed.One picture was received, and the review was made by an independent physician the results are shown below: ¿the description seems to suggest that the final stent did work.The single still image provided shows the stent being deployed in a tight curve with a returning segment (dipping mca).The observed landing zone is too short to allow opening of the distal portion of the stent resulting in all stent struts pointing straight ahead into the opposite wall.This could have been avoided by more distal opening of 50% of the stent with subsequent pull back to the target location, once the distal portion of the device opened.These curvatures are a known challenges for any stent, and only the kind of manipulation described, or taking a longer device and landing more distally, may help overcome this difficulty¿.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.The company is seeking this information through the event investigation.This report is being submitted pursuant to the provisions of 21 cfr, part 4.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.This is one of two products involved with the complaint and the associated manufacturer report numbers are 3008114965-2022-00574.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Updated sections on this medwatch: b4, b5, d9, g3, g6, h2, h3 and h10.Section b5: additional information received confirmed that the stents deployed but failed to expand.It is unknown the cause of the alleged product malfunctions.There was no severe vascular tortuosity at the target site.The system was used with the recommended microcatheter.The device nor the concomitant devices appear damaged.There was no resistance felt at any point of the procedure.It is unknown if it was necessary to remove the microcatheter.The product has been returned for evaluation and testing; however, the engineering evaluation has not been completed.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.Complaint conclusion: as reported by the field, during a stent assist coil embolization, a 4.5x28mm enterprise intracranial neurovascular stent 12 mm de tip (enc452812, 6920544) arrived at the target position and the physician tried to release it, but the stent could not be opened.The doctor changed to a new 4mmx30mm enterprise 2 intracranial neurovascular stent (encr403012, 6610181) but the same issue occurred.The third stent was switched to complete the surgery.There was no patient injury reported.Additional information received confirmed that the stents deployed but failed to expand.It is unknown the cause of the alleged product malfunctions.There was no severe vascular tortuosity at the target site.The systems were used with the recommended microcatheter.The device nor the concomitant devices appear damaged.There was no resistance felt at any point of the procedure.It is unknown if it was necessary to remove the microcatheter.One picture was received, and the review was made by an independent physician the results are shown below: the description seems to suggest that the final stent did work.The single still image provided shows the stent being deployed in a tight curve with a returning segment (dipping mca).The observed landing zone is too short to allow opening of the distal portion of the stent resulting in all stent struts pointing straight ahead into the opposite wall.This could have been avoided by more distal opening of 50% of the stent with subsequent pull back to the target location, once the distal portion of the device opened.These curvatures are a known challenges for any stent, and only the kind of manipulation described, or taking a longer device and landing more distally, may help overcome this difficulty.A non-sterile codman enterprise® vascular reconstruction device 4mm x 30mm was contained in the decontamination pouch.Upon receiving the device, a visual inspection was performed, and it was noted that the stent was returned already detached from the unit.The delivery wire and the introducer were also returned separated, and no damages were noted on them.The device was inspected under microscopic magnification, and the deployed stent was observed in good conditions, there was no structural damage (i.E., no broken struts, no kinks).No other damages were noted.Customer complaint regarding the stent not being able to open was not confirmed since the stent was found fully expanded during the analysis.With the limited information available, the root cause of the reported failure remains inconclusive; however, there is no indication that the issue reported in the complaint results from a defect inherently related to the device.A device history record (dhr) of the lot number 6610181 was performed and it indicated this product was final inspection tested at lake region medical and was determined to be acceptable.It should be noted that product failure could be caused by multiple factors.The instructions for use (ifu) do contain the following recommendations: ¿ do not apply undue force if resistance is encountered at any point during stent manipulation.Withdraw the unit and advance to a new one.¿ if resistance is felt while recapturing the stent, do not continue to recapture the device.Withdraw the infusion catheter slightly to unsheathe the stent (without exceeding the recapture limit), and then attempt to recapture the stent again.As part of the cerenovus quality process, all devices are manufactured, inspected, and released to approved specifications.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERPRISE2 4MMX30MM
Common Device Name
INTRACRANIAL NEUROVASCULAR STENT
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
Manufacturer (Section G)
CODMAN AND SHURTLEFF, INC
325 paramount dr
raynham MA 02767
Manufacturer Contact
michelleann garcia
31 technology dr
irvine, CA 92618
646591-7981
MDR Report Key15319436
Report Number3008114965-2022-00575
Device Sequence Number12053321
Product Code NJE
UDI-Device Identifier10886704075370
UDI-Public10886704075370
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
H60001
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/28/2024
Device Model NumberENF403012
Device Catalogue NumberENCR403012
Device Lot Number6610181
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/12/2022
Supplement Date Received by Manufacturer09/15/2022
09/28/2022
Initial Report FDA Received Date08/30/2022
Supplement Report FDA Received Date09/29/2022
10/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/21/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
EU 4.5X28MM STENT 12 MM DW TIP
Patient SexUnknown
-
-