• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL BCI WW1030 PULSE OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL BCI WW1030 PULSE OXIMETER Back to Search Results
Model Number WW1030EN
Device Problem Defective Alarm (1014)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Manufacturing device history record review was not performed because the results of the complaint investigation do not indicate a problem with the manufacturing of the device.A product sample was received for evaluation.Visual and functional testing were performed.The technician found out that the red light emitting diode (led) was damaged due to impact.The technician performed the functional test and reported issue of top red led does not flash when alarming was duplicated.The mainboard was also damaged.The technician replaced red led and mainboard to resolve the reported issue.The device operated as intended.The device passed all the functional and electrical testing.The root cause of the reported was determined to be related to the customer's impact to the device.Operator of device is unknown.No product information has been provided to date.This remediation mdr was generated under protocol b10009406, as a result of warning letter cms# (b)(4).
 
Event Description
It was reported that the top of the device does not flash when device alarm.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BCI WW1030 PULSE OXIMETER
Type of Device
OXIMETER
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15319998
MDR Text Key305150957
Report Number3012307300-2022-16689
Device Sequence Number1
Product Code DQA
UDI-Device Identifier30843418000191
UDI-Public30843418000191
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K083557
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberWW1030EN
Device Catalogue NumberWW1030EN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/18/2021
Was the Report Sent to FDA? No
Date Manufacturer Received02/16/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/25/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-