Model Number 402515
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/08/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The actual needles used at the time of the complaint were not returned for investigation.Unopened samples of the same lot number were received for investigation.Functional testing was done for patency by using a pin gauge.All samples passed.All samples were also tested by drawing up water through the needle and all performed as expected.The reported problem could not be confirmed based on the unused samples returned.A device history record (dhr) review found no discrepancies or anomalies relevant to the complaint.This remediation mdr was generated under protocol b10009406, as a result of warning letter cms# 617147.
 
Event or Problem Description
It was reported that while injecting the patient with depo, the needle became clogged half way through.A different needle of the same lot was tried and it also clogged.The user did a test with water and it clogged too.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HYPODERMIC NEEDLE-PRO EDGE SAFETY DEVICE
Common Device Name
NEEDLE, HYPODERMIC, SINGLE LUMEN
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
10 bowman dr.
keene NH 03431
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
10 bowman dr.
keene NH 03431
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15320272
Report Number3012307300-2022-16710
Device Sequence Number20134547
Product Code FMI
UDI-Device Identifier00351688070893
UDI-Public00351688070893
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K071785
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/12/2024
Device Model Number402515
Device Catalogue Number402515
Device Lot Number3912789
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/17/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/14/2022
Initial Report FDA Received Date08/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/02/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-