GLOBUS MEDICAL, INC. FORTIFY; FORTIFY Ø12MM CORE Back to Search Results
Model Number 151.052
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/19/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a fortify core has lost height 3 months post operatively.
 
Additional Manufacturer Narrative
The device was not available for evaluation as it remains in the patient.No determinations could be made as to the cause of the reported issue.
 
Event or Problem Description
It was reported that a revision surgery was needed to replace fortify core that lost height post operatively.
 
Additional Manufacturer Narrative
The device was returned for evaluation.Initial observation shows tissue is inside of the inserter hole touching the lock.The imaging provided shows tissue inside of the inserter hole.It's possible that the tissue may have caused the lock to move to the unlocked position.However, an exact cause of the reported issue could not be determined.The following sections have been updated: b4, d9, g6, h2, h3, h4, h6, h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORTIFY
Common Device Name
FORTIFY Ø12MM CORE
Manufacturer (Section D)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon PA 19403
Manufacturer (Section G)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon PA 19403
Manufacturer Contact
daniel paul
2560 general armistead avenue
audubon, PA 19403
6109301800
MDR Report Key15320274
Report Number3004142400-2022-00126
Device Sequence Number14753390
Product Code MQP
UDI-Device Identifier00889095042900
UDI-Public00889095042900
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K112756
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number151.052
Device Lot NumberBAY586BN
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/04/2022
Supplement Date Received by Manufacturer09/20/2022
Initial Report FDA Received Date08/30/2022
Supplement Report FDA Received Date10/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/05/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-