| Model Number |
EP8730H |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained via: this was only one incident ((b)(6) 2022), whereby two or three separate sutures from the same batch (# rmbdau) came away from the needle at the gentlest of pulls.This was all on one particular anastomosis during a cabg, which did mean the surgery was prolonged.Events reported via 2210968-2022-07072, 2210968-2022-07073.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a cabg procedure on (b)(6) 2022 and suture was used.During the procedure, the needles have broken away from the suture material.It was a control release needle.No adverse patient consequences were reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # = > pc-(b)(4).Date sent to the fda: 09/08/2022 h6 component code: g07002 - device not returned this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device, and no non-conformances were identified.Additional information was requested, and the following was obtained: they are not aware of how long in minutes the procedure was prolonged for.
|
| |
|
Search Alerts/Recalls
|