| Model Number |
8400 |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device had memory check error.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing device history record review was not performed because the results of the complaint investigation do not indicate a problem with the manufacturing of the device.A product sample was received for evaluation.Visual and functional testing were performed.No physical damage was observed.The technician performed the functional test and reported issue of memory check error was duplicated.The error was present on power up and would continue to reboot.The technician replaced the mainboard to resolve the reported issue.The device operated as intended.The device passed all the functional test.The root cause of reported issue could not be determined with the provided information.D5 is unknown.No product information has been provided to date.(b)(4).
|
| |
|