Model Number FV049T
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a control reservoir (part # fv049t) was implanted during an abdominal shunt surgical procedure performed on (b)(6) 2022.According to the complainant, the shunt system was believed to be blocked.Reportedly, a post operative re-examination found that the patient's cerebrospinal fluid was not able to be shunted out.Therefore, the patient underwent a revision procedure.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Same event as medwatch # 3004721439-2022-00258.
 
Additional Manufacturer Narrative
Manufacturing and quality control data the m.Blue® was manufactured by a qualified employee on january 2022.Deviations during assembly did not occur.The product was finally tested as article fv049t and released for packaging and sterilization and was sterilized by miethke.All tested parameters were assessed according to specifications.Visual inspection the following observations were made during the visual inspection: deposits on the reservoir.Permeability test.The test showed that the progav® 2.0 shunt system with the control reservoir are permeable.Results.Based on our investigation results, we cannot detect a blockage in control reservoir at time of investigation.The cause of the aforementioned functional impairment is not known to us at the time of the examination.We can exclude a defect at the time of release.The control reservoir met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTROL RESERVOIR SET WITH DISTAL CATHETER
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key15320475
Report Number3004721439-2022-00259
Device Sequence Number20210064
Product Code JXG
UDI-Device Identifier04041906131487
UDI-Public4041906131487
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K110206
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFV049T
Device Catalogue NumberFV049T
Device Lot Number20057344
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/15/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/02/2022
Initial Report FDA Received Date08/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/31/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age67 YR
Patient SexUnknown
-
-