Investigation: visual inspection: a deformation of the outer housing of the progav valve was observed through the visual inspection.The progav valve housing was subsequently measured and confirmed the presence of a deformation.The housing deformation measured at - 0.039 mm, outside the tolerance of 0 ± 0.02 mm.Permeability test: a permeability test has shown that the valve is permeable.Adjustment test: the progav was tested and is adjustable to all specified pressures.Klick force and brake function test: the brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Computer controlled test: to investigate the claim of over/under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in the horizontal position.The results show that the valve is operating within the accepted tolerance in the horizontal position.Internal inspection: after dismantling of the valve, deposits were found in progav.Result : based on our investigation results, we can determine deposits as well as a deformation of the outer housing.Neither the deposits nor the deformation of the outer housing had an impact to the technical properties at the time of the investigation.The cause of the aforementioned functional impairment is not known to us at the time of the examination.Proteins in the cerebrospinal fluid can influence the function temporarily and are known side effects in hydrocephalus therapy.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
|