BOSTON SCIENTIFIC CORPORATION VALITUDE CRT-P; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
|
Back to Search Results |
|
Model Number U125 |
Device Problems
Signal Artifact/Noise (1036); Failure to Capture (1081); High impedance (1291); Over-Sensing (1438); Pacing Problem (1439); High Capture Threshold (3266)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/15/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The local area field representative was contacted for additional information.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that this pacemaker, implanted with another manufacturer's right atrial (ra) lead, exhibited greater than two seconds of atrial pacing inhibition and there were stored signal artifact monitoring (sam) episodes that had disabled respiratory rate trend (rrt) due to noise on all channels.The patient was last seen in clinic on (b)(6) 2022 where it was noted that atrial pacing threshold was 3.5v@1ms and the output was set to 4.0v@1.0ms.It was suspected that the atrial pacing observed in the presenting and stored electrograms (egms) may be indicative of intermittent atrial loss of capture (loc).It was also noted that the patient had a barostim unit.Furthermore, there were a few spikes of high out-of-range pace impedance measurements greater than 2,000 ohms since february 2022.Technical services (ts) recommended further evaluation.This pacemaker remains in service.No adverse patient effects were reported.
|
|