• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION VALITUDE CRT-P; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION VALITUDE CRT-P; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P) Back to Search Results
Model Number U125
Device Problems Signal Artifact/Noise (1036); Failure to Capture (1081); High impedance (1291); Over-Sensing (1438); Pacing Problem (1439); High Capture Threshold (3266)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/15/2022
Event Type  malfunction  
Manufacturer Narrative
The local area field representative was contacted for additional information.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that this pacemaker, implanted with another manufacturer's right atrial (ra) lead, exhibited greater than two seconds of atrial pacing inhibition and there were stored signal artifact monitoring (sam) episodes that had disabled respiratory rate trend (rrt) due to noise on all channels.The patient was last seen in clinic on (b)(6) 2022 where it was noted that atrial pacing threshold was 3.5v@1ms and the output was set to 4.0v@1.0ms.It was suspected that the atrial pacing observed in the presenting and stored electrograms (egms) may be indicative of intermittent atrial loss of capture (loc).It was also noted that the patient had a barostim unit.Furthermore, there were a few spikes of high out-of-range pace impedance measurements greater than 2,000 ohms since february 2022.Technical services (ts) recommended further evaluation.This pacemaker remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALITUDE CRT-P
Type of Device
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key15320882
MDR Text Key302499182
Report Number2124215-2022-33250
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526559389
UDI-Public00802526559389
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030005/S113
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/20/2019
Device Model NumberU125
Device Catalogue NumberU125
Device Lot Number707291
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/23/2022
Initial Date FDA Received08/30/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/07/2017
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age67 YR
Patient SexFemale
-
-