• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAYENNE MEDICAL QUATTRO X SUTURE ANCHOR 5.5MM; FASTENER, FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAYENNE MEDICAL QUATTRO X SUTURE ANCHOR 5.5MM; FASTENER, FIXATION Back to Search Results
Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/05/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Cmp-(b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported during shoulder surgery, an anchor screw tip has broken during insertion in shoulder.An additional anchor was used to complete the procedure.No patient harm has been reported.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
Visual examination of the returned device identified the anchor is fractured just above the lead thread but the tip is held in place by the sutures threaded through it.There is some bio-residue on the anchor and on the tip of the inserter shaft.Medical records were not provided.Device history records were reviewed, and no discrepancies related to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUATTRO X SUTURE ANCHOR 5.5MM
Common Device Name
FASTENER, FIXATION
Manufacturer (Section D)
CAYENNE MEDICAL
16597 n 92nd street
scottsdale AZ 85260
Manufacturer (Section G)
CAYENNE MEDICAL
16597 n 92nd street
scottsdale AZ 85260
Manufacturer Contact
jennifer rapsavage
56 e. bell dr
warsaw, IN 46582
5745260384
MDR Report Key15320968
Report Number3006108336-2022-00027
Device Sequence Number5517960
Product Code MBI
UDI-Device Identifier00887868270277
UDI-Public(01)00887868270277(17)231211(10)69349-5
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K112814
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/11/2023
Device Model NumberN/A
Device Catalogue NumberCM-9255
Device Lot Number69349-5
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/05/2022
Supplement Date Received by Manufacturer11/16/2022
Initial Report FDA Received Date08/30/2022
Supplement Report FDA Received Date11/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/11/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
QUATTRO X DOUBLE LOADED PEEK ANCHOR.
Patient SexMale
-
-