• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ANGIOJET ULTRA 5000A; CATHETER, CORONARY, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ANGIOJET ULTRA 5000A; CATHETER, CORONARY, ATHERECTOMY Back to Search Results
Model Number 45038
Medical Device Problem Code Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/04/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the procedure was cancelled.An angiojet console was used for thrombectomy procedure.Upon preparation, the device showed an error 28.The device was removed and restarted but the error persisted.The procedure was canceled due to this event.There were no patient complications reported.
 
Event or Problem Description
It was reported that the procedure was cancelled.An angiojet console was used for thrombectomy procedure.Upon preparation, the device showed an error 28.The device was removed and restarted but the error persisted.The procedure was canceled due to this event.There were no patient complications reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr: angiojet barcode reader was received.A visual inspection was done on the barcode reader, and it passed visual inspection.The received barcode reader was installed onto the golden angiojet console.Then a functional test was done, and it failed the functional test.The console booted up to error code 28, and the complaint was confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIOJET ULTRA 5000A
Common Device Name
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15321032
Report Number2124215-2022-31993
Device Sequence Number14829201
Product Code MCX
UDI-Device Identifier08714729890584
UDI-Public08714729890584
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
P980037
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/07/2025
Device Model Number45038
Device Catalogue Number45038
Device Lot NumberOEMTRACE
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/04/2022
Supplement Date Received by Manufacturer01/13/2023
Initial Report FDA Received Date08/30/2022
Supplement Report FDA Received Date02/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/24/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-