STRYKER ORTHOPAEDICS-MAHWAH NO 8. TRIATHLON TS PLUS TIBIAL INSERT X3 POLY 16MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Model Number 5537-G-816
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Ambulation or Postural Difficulties (2544); Insufficient Information (4580)
Date of Event 08/09/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during the revision it was identified that the poly surrounding the post broke but the post remained in place.
 
Additional Manufacturer Narrative
An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
 
Additional Manufacturer Narrative
An event regarding crack/fracture involving a triathlon insert was reported.The event was confirmed via review of the provided photographs.Method & results: -device evaluation and results: the device was not returned for evaluation however photographs were provided for review.The photograph shows a recently explanted insert.Blood and biological matter is visible on the device.The post is fractured.The event is confirmed.-clinician review: no medical records were received for review with a clinical consultant.-device history review: could not be performed as lot code information was not provided.-complaint history review: could not be performed as lot code information was not provided.Conclusion: the device was not returned for evaluation however photographs were provided for review.The photograph shows a recently explanted insert.Blood and biological matter is visible on the device.The post is fractured.The exact cause of the event could not be confirmed as insufficient information was provided.Further information such as return of the device and primary operative reports are required to complete the investigation and determine a root cause.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
It was reported that during the revision it was identified that the poly surrounding the post broke but the post remained in place.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NO 8. TRIATHLON TS PLUS TIBIAL INSERT X3 POLY 16MM
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
joann ripoli
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15321133
Report Number0002249697-2022-01268
Device Sequence Number5230854
Product Code MBH
UDI-Device Identifier07613327027594
UDI-Public07613327027594
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number5537-G-816
Device Catalogue Number5537-G-816
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/09/2022
Supplement Date Received by Manufacturer11/07/2022
Initial Report FDA Received Date08/30/2022
Supplement Report FDA Received Date11/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
-
-