| Model Number |
2001M-PC |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Cardiac Arrest (1762)
|
| Date of Event |
07/05/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The sales representative reported on behalf of the customer that the 2001m-pc, padpro:ad;rtrans el/lw,pre-con, quantity of 3, were being used during a defibrillation due to cardiac arrest on (b)(6) 2022 when it was reported, ¿the padpro electrodes did not stick properly to the patient, why they had to be changed several times and delayed the defib.¿ further assessment questioning found that 3 pads were used on the same patient.There was no report of injury or prolonged hospitalization for the patient; however, the user did report that the scenario was ¿life threatening¿ because the pads did not work during a cardiac arrest defibrillation.This report is being raised on the basis of injury due to report of life-threatening scenario where pads did not stick to patient during use.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is not expected to be returned for evaluation and review.However, the complaint investigation is not complete at this time.A supplemental and final report will be filed following the completion of the complaint investigation.This issue will continue to be monitored through the complaint system to assure patient safety.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is not being returned and no photographic evidence was provided.Therefore, a device malfunction cannot be verified.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.(b)(4).Per the instructions for use, the user is advised to ensure that the patient¿s skin is clean and dry, clipping excess hair.Remove any substance from the skin (e.G., lotions, oils, etc.).1) tear open the pouch; 2) using the extended tab of the release liner, remove the release liner from the multifunction electrodes to expose conductive and adhesive areas.(arrow indicates peel location).This issue will continue to be monitored through the complaint system to assure patient safety.
|
| |
|
Event or Problem Description
|
|
The sales representative reported on behalf of the customer that the 2001m-pc, padpro:ad;rtrans el/lw,pre-con, quantity of 3, were being used during a defibrillation due to cardiac arrest on 5jul22 when it was reported, ¿the padpro electrodes did not stick properly to the patient, why they had to be changed several times and delayed the defib.¿.Further assessment questioning found that 3 pads were used on the same patient.There was no report of injury or prolonged hospitalization for the patient; however, the user did report that the scenario was ¿life threatening¿ because the pads did not work during a cardiac arrest defibrillation.This report is being raised on the basis of injury due to report of life-threatening scenario where pads did not stick to patient during use.
|
| |
|