SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX ENDOTRACHEAL TUBES POLAR; TUBE, TRACHEAL (W/WO CONNECTOR) Back to Search Results
Model Number 100/136/050
Medical Device Problem Codes Deflation Problem (1149); Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the pre-use check, the customer put air in the cuff and it was inflated after some delay.Also, similarly, there was a delay before the cuff was deflated.There is no additional information for this compliant.No patient injury reported.
 
Additional Manufacturer Narrative
A product sample was received and is awaiting evaluation, investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Additional Manufacturer Narrative
Device evaluation: three photos were included in agile attachments for evaluation; photos showed an inflation line occluded.One sample was returned for investigation.The returned sample was received in used condition without original packaging.The samples were visually inspected under normal conditions of illumination to detect any condition that could cause failure mode.An occlusion in the inflation line was detected.The returned sample was inflated using a syringe; the cuff would not inflate; the complaint was confirmed.The sample returned was connected to leak tester machine to identify any occlusion.Occlusion was detected.Based on the analysis conducted in the sample provided, cuff inflation/deflation failure mode was confirmed.The root cause was traced to the manufacturing process/assembly error for process not followed properly.For corrective action, customer complaint notification was made to production personnel to explain the importance of adherence or following in the procedure.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX ENDOTRACHEAL TUBES POLAR
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
3-10-15, hakata station east,
minneapolis, MN 55442
MDR Report Key15321262
Report Number3012307300-2022-16784
Device Sequence Number12142100
Product Code BTR
UDI-Device Identifier15019315021053
UDI-Public15019315021053
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K931735
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/10/2025
Device Model Number100/136/050
Device Catalogue Number100/136/050
Device Lot Number4062692
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/09/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/02/2022
Supplement Date Received by Manufacturer01/16/2023
Initial Report FDA Received Date08/30/2022
Supplement Report FDA Received Date06/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/12/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-