| Model Number |
DAT015 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
| Date of Event |
02/22/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
An ophthalmic technician reported that following an intraocular lens (iol) implant procedure, the patient was unhappy with results, unable to see intermediate or near vision.The iol was exchanged for another lens in a secondary procedure.Additional information was requested.
|
| |
|
Event or Problem Description
|
|
Additional information was received that, the patient was not hospitalized as a result of the event.The 21.0 diopter lens was exchanged for a 21.5 diopter lens due to poor near and intermediate vision.This does not meet criteria for reporting based on applicable medical device regulations.A lens exchange from one power to a different power is an attempt to adjust or fine tune the refractive outcome trying to achieve a desired target when the initial lens calculations did not achieve the intended outcome.There is no reported defect or fault with the lens.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on information received following submission of the initial report, this event does not meet criteria for reporting as a serious injury.The manufacturer internal reference number is: (b)(4).
|
| |
|