Model Number HDH05
Medical Device Problem Codes Material Separation (1562); Noise, Audible (3273)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/05/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch #: u94648.Additional information was requested and the following was obtained: the blade was broken outside the patient, and no pieces fell into the patient.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the hdh05 device was returned with the distal tip of the blade broken off and not returned.The remaining blade portion was scratched, and with evidence of contact with metal in or out of the operative field.In addition, it was noted the blade was loose from the shaft sheath.Upon visual inspection, it was noticed that the retention pin insulation was damaged.Damage to the pin coating could result in alert screens, such as "relaxed pressure in blade", ¿tighten assembly¿ or ¿blade error detected¿ followed by a ¿replace instrument¿ screen later in the procedure.Due to the device condition, no functional test could be performed.No conclusion could be reached as to what caused this issue.It is possible that the reported noise issue could have been caused by metal to metal contact once the blade was damaged.As part of the quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances related to the reported complaint condition were identified.
 
Event or Problem Description
It was reported that, during an ob-gyn procedure, an unexpected metallic sound was heard.The blade was broken.No pieces fell into the patient.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC SCALPEL 5MM HD SHEATH HOOK TIP
Common Device Name
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
*  
329348013
MDR Report Key15321388
Report Number3005075853-2022-05696
Device Sequence Number5230356
Product Code LFL
UDI-Device Identifier10705036002109
UDI-Public10705036002109
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K990362
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2025
Device Model NumberHDH05
Device Catalogue NumberHDH05
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/05/2022
Initial Report FDA Received Date08/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/27/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GENERATO
Patient SexUnknown
-
-