MAUDE Adverse Event Report: TELEFLEX MEDICAL MDK VI MF 0 TC43/HR26 48"; SUTURE, SURGICAL, ABSORBABLE,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL MDK VI MF 0 TC43/HR26 48"; SUTURE, SURGICAL, ABSORBABLE, Back to Search Results
Model Number IPN000864
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/20/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).A visual, dimensional, or functional inspection of the product involved in the complaint could not be conducted since the product was not returned.A dhr review could not be conducted since the lot number of the reported product code was not provided.No corrective action can be implemented due to the lack of batch number to perform a proper investigation to determine a root cause.The customer complaint cannot be confirmed based only on the information provided, to perform an investigation and determine the source of defect reported it is necessary to evaluate the sample involved on this complaint.An attempt to duplicate the failure mode was made but at the time there is no inventory of the involved product code available at the facility nor is being manufactured at the time.If the sample becomes available this investigation will be updated with the evaluation results.
 
Event or Problem Description
Physician says that when he fires the suture to deploy through the ligament, the bullet on the end of the suture gets stuck.He attempted a second time and the device misfired.Last attempt the bullet came off suture.Physician opened a new capio slim and completed the procedure with no issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MDK VI MF 0 TC43/HR26 48"
Common Device Name
SUTURE, SURGICAL, ABSORBABLE,
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
parque industrial finsa
nuevo laredo 88275
MX   88275
Manufacturer Contact
effie jefferson
3015 carrington mill blvd
morrisville 27560
9194332672
MDR Report Key15321801
Report Number3004365956-2022-00060
Device Sequence Number5518489
Product Code NEW
UDI-Device Identifier14026704603920
UDI-Public14026704603920
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K030212
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIPN000864
Device Catalogue Number833-137
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/16/2022
Initial Report FDA Received Date08/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-