• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 LUBRI-SIL® I.C. COMPLETE CARE® ALL-SILICONE FOLEY CATHETER TRAY AND URINE METER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 LUBRI-SIL® I.C. COMPLETE CARE® ALL-SILICONE FOLEY CATHETER TRAY AND URINE METER Back to Search Results
Model Number 303416A
Medical Device Problem Code Component Misassembled (4004)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that there was no lubricant in the lubricant jelly syringe.It was also stated that the syringe that was supposed to contain saline did not have a plunger and therefore there was no saline.
 
Event or Problem Description
It was reported that there was no lubricant in the lubricant jelly syringe.It was also stated that the syringe that was supposed to contain saline did not have a plunger and therefore there was no saline.Per additional information received via sample form on (b)(6) 2022, it was stated that when nurse was prepping to insert catheter, upon inspection of the kit prior to insertion, it was noted that there was no lube in the lube syringe and sterile saline syringe had no plunger.Another tray was opened and used without any issue.
 
Additional Manufacturer Narrative
The reported event is confirmed manufacturing related.Visual evaluation noted received 1 foley tray system without packaging.Visual inspection noted lubricant residue in the lubricant jelly syringe.Also noted the sterile water syringe did not have a plunger and therefore there was no saline residue in the syringe.Also, noted that the plunger was missing from the tray.Therefore, product does not meet specifications according to inspection procedure which states, "no missing components allowed." and "water fill volume: 9.6 + 0.4/-0.6cc."although an exact root cause could not be determined a potential root cause is incorrect operation.The lot number is unknown; therefore, the device history record could not be reviewed.Labelling review is not required as labelling would not have prevented the reported event.Correction: d, g, h h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUBRI-SIL® I.C. COMPLETE CARE® ALL-SILICONE FOLEY CATHETER TRAY AND URINE METER
Common Device Name
ALL-SILICONE FOLEY CATHETER TRAY
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key15321959
Report Number1018233-2022-06794
Device Sequence Number5124858
Product Code EZL
UDI-Device Identifier00801741024726
UDI-Public(01)00801741024726
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K040504
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number303416A
Device Catalogue Number303416A
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/19/2022
Supplement Date Received by Manufacturer02/09/2023
Initial Report FDA Received Date08/30/2022
Supplement Report FDA Received Date02/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-