• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. SINGLE USE INJECTOR NM600/610; INJECTOR AND SHEATHSET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. SINGLE USE INJECTOR NM600/610; INJECTOR AND SHEATHSET Back to Search Results
Model Number NM-610U-0426
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/03/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is not returned, as such a definitive root cause of the reported complaint cannot be determined at this time.This event is under investigation.A supplemental report will be submitted upon completion of the investigation or upon receiving additional information.
 
Event or Problem Description
As reported for this event by the customer, during an unknown procedure the customer had difficulty in opening the sealed package of the device.This device is sealed with an additional seal package.Since the customer had to forcibly open the package, there is a chance of contamination by contact with an unclean area.Customer is recommending that the package seal be improved for ease of opening.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.H4: the subject device was manufactured in may 2022.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The legal manufacturer's final investigation confirmed that the sterile packages that are additionally heat-sealed have a higher seal strength than those that are not additionally heat-sealed.Therefore, the sterile packages with additional heat sealing are difficult to open.The legal manufacturer has determined there is no potential for this malfunction to cause or contribute to an adverse event.
 
Additional Manufacturer Narrative
Additional information has been received for this event.This supplemental report is being submitted to provide this information.Please see the updates in sections: b5, g3, g6, h2, and h10.Patient information is not known.
 
Event or Problem Description
Though the customer complained about the difficulty in opening the device package seal, the customer was able to open the packaging and use the device in the procedure.There was no harm or adverse impact on the patient.The procedure details are not known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE INJECTOR NM600/610
Common Device Name
INJECTOR AND SHEATHSET
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer Contact
masaharu hirose
2-248-1 okkonoki
kuroishi-shi, aomori 036-0-357
JA   036-0357
426422891
MDR Report Key15322163
Report Number9614641-2022-00230
Device Sequence Number12062902
Product Code FBK
UDI-Device Identifier04953170422874
UDI-Public04953170422874
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
K153625
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNM-610U-0426
Device Lot Number25K
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/03/2022
Supplement Date Received by Manufacturer09/06/2022
10/03/2022
Initial Report FDA Received Date08/30/2022
Supplement Report FDA Received Date09/20/2022
10/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-