Model Number 100/166/075
Medical Device Problem Code Material Invagination (1336)
Health Effect - Clinical Code Decreased Respiratory Rate (2485)
Date of Event 07/22/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that after patient intubation a ventilation issue was observed.After examination via fibroscopy, visible folding of the intubation tube was found to be preventing ventilation.Tube was in place for fourteen days.Removal of the tube with guide and reintubation with a new tube resolved the issue.Additional information was received on 4-aug-2022 stating no clinical consequence to the patient was observed.Premature tube change had to take place.Additional information received on 16-aug-2022 stating the tube was bent around and between the base of the tongue and the pharynx about in the middle of the tubes.Incident occurred with three (3) patients after fourteen (14) days (patient 1), nine (9) days (patient 2) and one (1) day (patient 3).Intubation was not difficult for all three patients and under direct laryngoscopy.The patients were in intensive care unit.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when required.
 
Additional Manufacturer Narrative
H10: a device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample and four photos were received for evaluation.Visual inspection found the photos showed an inflation line occluded.No other analysis was performed.The root cause of the reported issue was found to be supplied component fault.Customer complaint notification was made to production personnel to explain the importance to adherence or following in the procedure by quality engineer on 20/sep/2022.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX TRACHEAL TUBE SILICONIZED PVC SOFT SEAL CUF
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15322257
Report Number3012307300-2022-16817
Device Sequence Number5206383
Product Code BTR
UDI-Device Identifier15019315054099
UDI-Public15019315054099
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K790312
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/20/2026
Device Model Number100/166/075
Device Catalogue Number100/166/075
Device Lot Number4150640
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/02/2022
Supplement Date Received by Manufacturer03/23/2023
Initial Report FDA Received Date08/30/2022
Supplement Report FDA Received Date04/04/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/16/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-