Model Number 8400D
Medical Device Problem Codes Device Alarm System (1012); Failure to Power Up (1476)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturing device history record review was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.No causes or potential causes of the customer's reported problem were found during the review of service and repair records.A product sample was received for evaluation.Visual and functional testing were performed.Visual inspection found loose speaker and dislodged pump.Functional testing could not confirmed reported problem.The root cause of the issue could not be determined; however, there is evidence of physical impact.This remediation mdr was generated under protocol b10009406, as a result of warning letter cms# (b)(4).
 
Event or Problem Description
It was reported that "c-sum" error was displayed.Device would not power up.Kept flashing new "co2 sensor".No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE
Common Device Name
CAPNOCHECK II, MODEL 8400
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15322260
Report Number3012307300-2022-16818
Device Sequence Number5888335
Product Code CCK
UDI-Device Identifier10610586036736
UDI-Public10610586036736
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K991086
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8400D
Device Catalogue Number8400D
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2021
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2022
Initial Report FDA Received Date08/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/10/2012
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-