Model Number TGMR373720
Medical Device Problem Code Failure to Seal (4070)
Health Effect - Clinical Code Vascular Dissection (3160)
Date of Event 08/12/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Patient medical history includes but is not limited to: thoracic dissection, coronary artery disease, hyperlipidemia and svt.Concomitant medical products:: patient medications include but are not limited to: tylenol, albuterol, amiodarone and melatonin.(b)(4).W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
On (b)(6) 2020, this patient underwent treatment of a type b thoracic dissection and was implanted with gore® tag® conformable thoracic stent grafts.The patient tolerated the procedure.On (b)(6) 2022, the physician reported degeneration of the proximal seal of the most proximal gore® tag® conformable thoracic stent graft, and an extension of the dissection proximally on the inner curve at the level of the left subclavian artery (lsa).The physician believes the extension of the dissection was caused/related to the anastamosis in the arch area near the lsa.On (b)(6) 2022, the patient underwent reintervention and the extension of the dissection gore® tag® thoracic branch endoprostheses.The patient tolerated the procedure with great results.
 
Additional Manufacturer Narrative
Images were provided to gore for evaluation and showed the following: the aortic diameter just distal to the lsa appears to be 34.6mm.There is a tear ~15mm ¿ 16.7mm proximal to the distal border of the lsa.This finding is consistent with the description summary.The proximal tear is on the inner curve.The false lumen is being perfused from the tear on the inner curve and communicates with contrast on the outer curve.The dissection extends to the celiac artery/end of scan.
 
Additional Manufacturer Narrative
H6: corrected conclusion code.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® TAG® CONFORMABLE THORACIC STENT GRAFT WITH ACTIVE CONTROL SYSTEM
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL WOODY SPRINGS B/P
3450 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
laura crawford
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key15322611
Report Number2017233-2022-03277
Device Sequence Number5518009
Product Code MIH
UDI-Device Identifier00733132646821
UDI-Public00733132646821
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P040043
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/03/2022
Device Model NumberTGMR373720
Device Catalogue NumberTGMR373720
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/12/2022
Supplement Date Received by Manufacturer08/12/2022
08/12/2022
Initial Report FDA Received Date08/30/2022
Supplement Report FDA Received Date09/08/2022
12/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/04/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention; Other;
Patient SexMale
Patient Weight102 KG
-
-