• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILLIAM COOK EUROPE GUNTHER TULIP NAVALIGN FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILLIAM COOK EUROPE GUNTHER TULIP NAVALIGN FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Model Number G52917
Medical Device Problem Code Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Serious Injury
Event or Problem Description
Description of event according to initial reporter: the white plastic hub that houses the safety and releases mechanism was not aligned properly.They stated it was a very small misalignment.They realigned it and continued with the procedure.When trying to deploy, the filter would not disengage from the release mechanism.They then tried to re-sheath it and couldn¿t cover the hook on the filter.On the way out the filter got snagged and released in the femoral vein.The filter was off of the device but the feet were not deployed.The filter then had to be snared from a jugular vein approach.Patient outcome: no harm came to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUNTHER TULIP NAVALIGN FEMORAL VENA CAVA FILTER SET
Common Device Name
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
MDR Report Key15322848
Report Number3005580113-2022-00099
Device Sequence Number12129463
Product Code DTK
UDI-Device Identifier10827002529172
UDI-Public(01)10827002529172(17)241125(10)E4191246
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/30/2022
Report Date (Section F) 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/25/2024
Device Model NumberG52917
Device Catalogue NumberIGTCFS-65-1-FEM-TULIP
Device Lot NumberE4191246
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date08/09/2022
Device Age9 YR
Event Location Hospital
Date Report to Manufacturer08/30/2022
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/30/2022
Date Device Manufactured11/25/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-