SMITH & NEPHEW, INC. SCREW BIOSURE REGENESORB 7MM X 25MM; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 72204395
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Device Embedded In Tissue or Plaque (3165); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/10/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference case: (b)(4).
 
Event or Problem Description
It was reported that during the femoral fixation procedure of the graft performed with the biosure regenesorb screw, after inserting more than half of the screw according to the instructions for use and turning the screwdriver, the head of the screw broke.This head was removed from the joint with the help of forceps.The graft was tightened and it was visualized that the graft does not move so there was no need to remove the screw and replace it with another one.The procedure was completed with non-significant surgical delay and was finished with the same smith and nephew device.No further complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4).H10 h3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A review of the polymer found that the storage requirements for the material are specified, and the material must comply with provided percentage composition specifications as measured by ash test.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.This case reports that the head of the biosure regenesorb screw, broke during the procedure and while the head was retrieved, the bottom portion of the screw piece was left in the patient.No clinically relevant supporting documentation was provided for inclusion in this medical investigation.Based on the information provided no clinical factors were found which would have contributed to the reported breakage.The device instruction for use does note that ¿excessive force should not be placed on the screw, bone, or delivery instruments.¿ the biosure screw is made of a biocomposite material which is intended for implantation to allow for bone ingrowth.It is unlikely that retained piece of biosure screw will migrate as it was noted that, ¿the graft was tightened and it was visualized that the graft does not move so there was no need to remove the screw and replace it with another one.¿ there is potential risk for tissue inflammation and/or pain.No further clinical/medical assessment is warranted at this time.There was no relationship found between the device and the reported event.The complaint was not confirmed.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCREW BIOSURE REGENESORB 7MM X 25MM
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15323428
Report Number1219602-2022-01282
Device Sequence Number5518542
Product Code MAI
UDI-Device Identifier00885554036718
UDI-Public00885554036718
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K142948
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/18/2024
Device Model Number72204395
Device Catalogue Number72204395
Device Lot Number50960623
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/10/2022
Supplement Date Received by Manufacturer10/03/2022
Initial Report FDA Received Date08/30/2022
Supplement Report FDA Received Date10/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/18/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age41 YR
Patient SexFemale
-
-