Model Number 26173HA
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The item in question was sent to the manufacturer.The evaluation is anticipated, but not yet begun.The investigation will be performed by a designated karl storz employee.
 
Event or Problem Description
As per a manufacturer incident report (b)(4) received from the factory in germany: it was reported that the physician used the device for suture during an ovarian cyst removing surgery on (b)(6) 2022.The tip of the needle holder broke under normal use and fell into the patient's abdomen.In order to retrieve the fallen part, the user prolonged the anesthesia time and used c-arm x-ray to locate the broken jaw.The operation time was extended for more than two hours.The detached part was found successfully.In order to make sure the patient's overall health and safety, the hospitalization was prolonged for 3 days.There were no long term injuries for the patient.
 
Additional Manufacturer Narrative
The corresponding manufacturer's complaint reference number for this case is (b)(4).Second supplemental report was initiated to correct section g3.
 
Additional Manufacturer Narrative
The corresponding manufacturer's complaint reference number for this case is (b)(4).The affected 26173ha needle holder was returned to the manufacturer on 2022-08-29 and the investigation was concluded on 2022-09-29.Upon initial inspection of the product, it could be confirmed that it is damaged at the distal end.The fixed jaw part has broken off from the needle holder.During further investigation, it was found that the axle pin which holds the movable jaw in the fixed jaw is visible from the outside.Normally it is not visible here, since it is welded from the outside, but in this case, the welding seam was broken.This indicates that a high force must have been applied for the welding seam to break.In addition, there are impurities and discolorations visible on the needle holder which have led to corrosion around the needle holder where the breakage happened.Therefore, it can be concluded that the material was weakened due to corrosion, which contributed to the breakage of the fixed jaw part when the high force was applied.Consequently, a combination of corrosion and overloading of the jaw part is concluded as the most probable root cause.Based on the impurities and discolorations, it can furthermore be concluded that the reason for the corrosion was that the needle holder was not adequately cleaned and properly dried after reprocessing, allowing residual moisture to accumulate.The ifu includes relevant information regarding the cleaning of the device, and it also warns against overloading by exerting too much force.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEEDLE HOLDER
Common Device Name
NEEDLE HOLDER, REUSABLE
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM  78532
Manufacturer (Section G)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM   78532
Manufacturer Contact
anja fair
2151 e grand ave
el segundo, CA 90245-5017
4242188738
MDR Report Key15323988
Report Number9610617-2022-00258
Device Sequence Number14832378
Product Code GCJ
UDI-Device Identifier04048551387405
UDI-Public4048551387405
Combination Product (Y/N)N
Initial Reporter CountryTW
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number26173HA
Device Catalogue Number26173HA
Device Lot NumberUP01
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer09/29/2022
09/29/2022
Initial Report FDA Received Date08/30/2022
Supplement Report FDA Received Date10/06/2022
10/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-