Model Number 33650001
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/04/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device is not available for evaluation as it remains in the patient.A review of the device history is not possible because the lot number was not communicated.Should additional information become available, it will be provided on a supplemental report.
 
Event or Problem Description
The patient underwent a total ankle replacement surgery.Allegedly, the patient may need to undergo a revision surgery due to reasons that were not available at the time of this report.
 
Additional Manufacturer Narrative
Please note the corrections to the h6 results code: the reported event could be confirmed since images of ct scans were provided and matches the alleged failure mode.There are many clinical factors that can affect the results of any surgery, such as surgical technique, pre-operative and post-operative care, the implant, patient pathology and daily activity.A medical professional reviewed the received information and noted the following: ¿the ct-scan shows loosening and subsidence of both tibial and talar components.The pe-liner looks intact and at the right place, as far as can be assessed on a ct-scan.There are no signs of any device related problems.Most likely the loosening and subsidence of both tibial and talar components relate to poor bone quality and poor joint mechanics and/or low-grade infection, which can only be assessed with further clinical information.¿ more detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
The patient underwent a total ankle replacement surgery.Allegedly, the patient may need to undergo a revision surgery due to reasons that were not available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITY TIBIAL TRAY SIZE 1 TOTAL ANKLE SYSTEM
Common Device Name
PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY INC
1023 cherry rd
memphis TN 38117
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15324261
Report Number3010667733-2022-00289
Device Sequence Number5505251
Product Code HSN
UDI-Device Identifier00840420122483
UDI-Public00840420122483
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K140749
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number33650001
Device Catalogue Number33650001
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/04/2022
Supplement Date Received by Manufacturer09/28/2022
Initial Report FDA Received Date08/30/2022
Supplement Report FDA Received Date10/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;