• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CHRISTOPH MIETHKE GMBH & CO KG M.BLUE 15 VALVE; HYDROCEPHALUS MANAGEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CHRISTOPH MIETHKE GMBH & CO KG M.BLUE 15 VALVE; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FX803T
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation: visual inspection: deposits in the inlet spout.Permeability test: a permeability test has shown that the valve is permeable.Adjustment test: the m.Blue valve was tested and is adjustable to all specified pressures.Braking force and brake function test: the brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Computer controlled test: to investigate the claim of over/under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The results show that the valve operates within the accepted tolerance in both positions.Internal inspection: after dismantling of the valve, deposits were found in m.Blue.Results: for the sake of thoroughness and transparency, we would like to point out that an adjustment test and permeability test was already carried out after the revision at the hospital.Based on our investigation results, we can determine deposits.The deposits had no effect to the technical properties at the time of the investigation.The cause of the aforementioned functional impairment is not known to us at the time of the examination.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Event or Problem Description
It was reported that a m.Blue (part # fx803t) was implanted during a procedure performed on an unknown date.According to the complainant, the shunt system was believed to be operated in underdrainage.The patient underwent a revision procedure performed on an unknown date.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 8 years, height: 129 centimeters (cm), weight: 28 kilograms (kg), gender: male.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M.BLUE 15 VALVE
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key15324282
Report Number3004721439-2022-00267
Device Sequence Number5505747
Product Code JXG
UDI-Device Identifier04041906504427
UDI-Public4041906504427
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K192266
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFX803T
Device Catalogue NumberFX803T
Device Lot Number20056242
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/09/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/09/2022
Initial Report FDA Received Date08/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age8 YR
Patient SexMale
Patient Weight28 KG
-
-