| Catalog Number |
09556494001 |
| Medical Device Problem Code |
False Negative Result (1225)
|
| Health Effect - Clinical Code |
Bacterial Infection (1735)
|
| Date of Event |
07/20/2022
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
Analysis: internal review of qc release data for affected lot 91246383 confirms the consumable lot successfully met the established qc release specifications.Review of the eplex bcid-gn panel test run information demonstrates that the internal controls were valid.No run malfunction was observed and the eplex instrument (serial #1922100442) was working within design specifications.Conclusion: while a discrepant result was reported by the customer, no quality and/or performance issues have been identified for the affected eplex bcid-gn panel lot.Genmark¿s investigation is ongoing, as the sample has been requested from the customer.The eplex bcid-gn panel was appropriately used according to the product's intended use, which describes the assay as an aid in the diagnosis of bloodstream infection when used in conjunction with other clinical information.The results from the eplex bcid-gn panel are intended to be interpreted in conjunction with gram stain results and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2022, genmark was advised of unexpected negative results for pseudomonas aeruginosa on the eplex bcid-gn panel tested on (b)(6) 2022.On (b)(6) 2022, during communication related to the investigation genmark was advised that patient was deceased.The customer was not able to confirm whether the result of the eplex contributed to the death of the patient and referred to the patient history of liver failure and previous admissions for pseudomonas cholangitis/sepsis.Genmark was advised that the patient had been on 2 grams of iv meropenem every 12 hours for two weeks prior to being admission on (b)(6) 2022 with weakness, fevers, and chills.Background: an initial set of blood culture bottles collected on (b)(6) 2022 was not tested on eplex and showed only stenotrophomonas growing in culture.A second set of blood culture bottles were collected on (b)(6) 2022 and this time stenotrophomonas and pseudomonas were seen in culture.An aliquot from this set of blood culture bottles was tested with the eplex bcid-gn on (b)(6) 2022 where only stenotrophomonas maltophilia was detected.Testing a second aliquot from the same bottle with the eplex bcid-gn again resulted in only stenotrophomonas maltophilia being detected.A third set of blood culture bottles were collected on (b)(6) 2022 and only stenotrophomonas was seen in culture.Data was analyzed per internal procedures and confirmed robust internal controls were generated for both eplex bcid-gn runs suggesting the consumables performed as expected.
|
| |
|
Search Alerts/Recalls
|