| Model Number |
03P6822 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.All available patient information was included.
|
| |
|
Event or Problem Description
|
|
The customer stated that a falsely elevated architect magnesium result of 6.79 mg/dl was generated for a patient sample (b)(6) that retested at 2.0 and 2.1 mg/dl.Reference range: 1.9 - 2.8 mg/dl no impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Trending review determined no adverse trend for the issue for the product.Historical complaint review determined there is normal complaint activity for the lot number.Labeling was reviewed and found to adequately address the issue under review.Worldwide data from abbottlink was reviewed and determined that patient median result for the lot is comparable with other lots in the field and within established baselines confirming no systemic issue for the lot.Based on the investigation, no deficiency for lot number 14442un22 was identified.This follow up includes a correction to include the codes in section h6 for component code, type of investigation, investigation findings and investigation conclusions that were not included in the initial report.
|
| |
|
Search Alerts/Recalls
|