Blank fields on this form indicate the information is unchanged, unknown, or unavailable.Investigation ¿ evaluation a representative of china medical university hospital (taiwan) informed cook on (b)(6) 2022 of an issue with a yueh centesis disposable catheter needle (rpn: dtvn-5.0-19-15.0-yueh; lot #: 13391496).It was reported that the device broke at the hub during a percutaneous abscess drainage procedure on a 73-year-old male patient.The user opened a new device to complete the procedure successfully.The patient did not experience any adverse effects due to this occurrence.Reviews of the complaint history, device history record (dhr), and quality control procedures, as well as a visual inspection of the returned device, were conducted during the investigation.One used device was received.The hub appears to have snapped into two pieces.Additionally, a document-based investigation evaluation was performed.In response to this incident, cook completed a review of the product device master record (dmr) and concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.Cook also reviewed the device history record (dhr).The dhr for lot 13391496 found no additional complaints or nonconformances.Cook requested that the supplier investigate this occurrence.The device history record showed that the affected lot was released accomplishing all quality standard requirements.There was no deviation and no non-conformances reported during the manufacturing of this product.A review of the internal manufacturing device record, raw material history files and sub assembly catheter hub part number and the reported lot number was conducted, and no recorded quality problems or rejections related to this incident were found.It was confirmed that procedural and functional requirements needed for its release were met.Based on the dmr, dhr, device evaluation, and supplier investigation, there is no indication the complaint device was manufactured out of specification.There is no evidence of nonconforming material in house or in the field.Cook reviewed product labeling.This device is not currently packaged with instructions for use.Based on the information provided, inspection of the returned device, and the results of the investigation, cook could not establish a cause for this event.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|