Manufacturer Narrative
|
A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that, the cardiosave intra-aortic balloon pump (iabp) unit had a problem with its fiber optic port.
|
|
Event Description
|
It was reported that, the cardiosave intra-aortic balloon pump (iabp) unit had a problem with its fiber optic port.There was no patient incident reported.
|
|
Manufacturer Narrative
|
Updated data: b4,g3,g6,h2,h10,h11.Corrected data: b5,e3,h6 (clinical & impact).
|
|
Manufacturer Narrative
|
Updated fields: b4, d9, e1 (site country), g3, g6, h2, h3, h4, h6 (type of investigation, investigation findings, component codes and investigation conclusions).A getinge field service engineer (fse) was dispatched to investigate.The fse evaluated the cardiosave intra-aortic balloon pump (iabp) unit and found that the plastic connector that the fiber plugs into was cracked and broken.It still tested ok but the connection was not very stable.To fix the issue, the fse replaced the cbl assy,fo sensor extension (0012-00-1562).The fse performed functional and safety checks to meet factory specifications.The iabp was then released to the customer and cleared for clinical service.
|
|