Model Number 0998-00-0800-53
Device Problem Failure to Sense (1559)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/18/2022
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that, the cardiosave intra-aortic balloon pump (iabp) unit had a problem with its fiber optic port.
 
Event Description
It was reported that, the cardiosave intra-aortic balloon pump (iabp) unit had a problem with its fiber optic port.There was no patient incident reported.
 
Manufacturer Narrative
Updated data: b4,g3,g6,h2,h10,h11.Corrected data: b5,e3,h6 (clinical & impact).
 
Manufacturer Narrative
Updated fields: b4, d9, e1 (site country), g3, g6, h2, h3, h4, h6 (type of investigation, investigation findings, component codes and investigation conclusions).A getinge field service engineer (fse) was dispatched to investigate.The fse evaluated the cardiosave intra-aortic balloon pump (iabp) unit and found that the plastic connector that the fiber plugs into was cracked and broken.It still tested ok but the connection was not very stable.To fix the issue, the fse replaced the cbl assy,fo sensor extension (0012-00-1562).The fse performed functional and safety checks to meet factory specifications.The iabp was then released to the customer and cleared for clinical service.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key15324609
MDR Text Key302395650
Report Number2249723-2022-02221
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date 01/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/16/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2015
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-