| Model Number |
RETROVERSION HNDL |
| Medical Device Problem Code |
Device Damaged by Another Device (2915)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/02/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
It was reported that during a surgical procedure, when removing a trial stem using the stem extractor, the surgeon accidentally struck the retroversion rod, causing the screw thread to snap off within the stem extractor tool.No fragments, pieces or parts fell into patient wound.Patient was last known to be in stable condition following the event.The device is available for return.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on historical complaint investigations, the returned device, and the reported event, the broken retroversion handle reported was likely the result of the surgeon accidently impacting the retroversion handle while removing a trial stem, leading to ultimate failure of the retroversion handle.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon reassessment of the reported event, it was determined that the event did not involve a death or serious injury nor would be likely to cause or contribute to a death or serious injury if the reported malfunction were to recur.As a result, this complaint event is no longer considered reportable by exactech and this report may be disregarded.
|
| |
|