Trackwise # (b)(4).Updated section: b4, g4, g7, h2, h3, h6, h10.The device was returned to the factory for evaluation on 08/17/2022.An investigation was conducted on (b)(6) 2022.A visual inspection was conducted.Signs of clinical use and evidence of blood were observed on the harvesting tool, cannula handle, syringe attached to the scope wash tubing, cannula shaft, c-ring, and jaws.Charred tissue was also observed on the jaws of the harvesting device.No visual defects were observed on the c-ring of the cannula or heater wire and clear silicone insulation of the jaws.A pre-cautery test was performed per the instruction for use (ifu) with a reference cable, adapter, and reference power supply vh-3010 at level 3.0.The device passed the pre-cautery test; it produced a normal amount of steam and heat during ten (10) 3-second activations and shut off when the toggle was released.No excessive smoke and/or steam were observed during the testing.There was no smoke observed coming from the harvesting handle near the toggle switch.Based on the returned condition of the device as well as the evaluation results, the reported failure "environmental particulates; excessive smoke" was not confirmed.The lot history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
|