| Model Number |
637-202 |
| Medical Device Problem Codes |
Device Alarm System (1012); Excess Flow or Over-Infusion (1311); Insufficient Flow or Under Infusion (2182)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical inc.Internal report: (b)(4).The complaint is under evaluation.A follow-up report will be provided after the examination results are available.
|
| |
|
Event or Problem Description
|
|
As reported by the user facility: "malfunction code 108 ff1.6.Infusions running slow.(when i turned it on there was no malfunction code present, however the pump had been turned off and unplugged for".
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).The reported issue was present during investigation.Incoming pump inoperable and alarming.Replace membrane keypad #34520635 and confirmed correct delivery accuracy.Perform preventive maintenance service.Log review confirms alarms 016 and 108.
|
| |
|