Model Number 7132
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: used the first day of the month of aware date, as event date was not provided.
 
Event or Problem Description
It was reported that balloon rupture occurred.The target lesion was located in the left anterior descending artery.A 2.50mm x 12mm emerge balloon catheter was selected for use.However, during preparation, the balloon ruptured.The procedure was completed with a different device.There were no patient complications reported.
 
Event or Problem Description
It was reported that balloon rupture occurred.The target lesion was located in the left anterior descending artery.A 2.50mm x 12mm emerge balloon catheter was selected for use.However, during preparation, the balloon ruptured.The procedure was completed with a different device.There were no patient complications reported.
 
Additional Manufacturer Narrative
B3 date of event: used the first day of the month of aware date, as event date was not provided.Device evaluated by mfr.: returned product consisted of an incomplete device return of an emerge mr balloon catheter.The device was visually and microscopically examined.At 51.5cm from the strain relief the hypotube was separated, which was also the length of the device returned.The separated end was ovaled in shape indicating the device was kinked prior to separation.There were numerous kinks to the hypotube.No further damage was found to the returned portion of the device which included the manifold, strain relief, and a portion of the hypotube.Product analysis could not confirm the reported burst as the balloon portion of the device failed to return for further analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMERGE
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15325514
Report Number2124215-2022-31428
Device Sequence Number20423397
Product Code LOX
UDI-Device Identifier08714729806103
UDI-Public08714729806103
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K113220
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/23/2024
Device Model Number7132
Device Catalogue Number7132
Device Lot Number0029308166
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/29/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/12/2022
Supplement Date Received by Manufacturer09/28/2022
Initial Report FDA Received Date08/30/2022
Supplement Report FDA Received Date09/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/28/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-