| Model Number |
MA60AC |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported by pharmacist that during posterior vitrectomy surgery and phacoemulsification surgery, fissure was noted in the optic, the lens was changed and implanted without problems.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
The lens was returned positioned incorrectly in the lens case.Solution was dried on the lens.The optic had scrape marks and cracks on the posterior side of the lens.Associated products were not provided.It is unknown if qualified products were used.The company model is only qualified for use in the company cartridges.The reported damage was observed.All lenses are 100% inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met company release criteria.Based on our observation, and the review of manufacturing records, it can be reasonably concluded that the damage is not manufacturing related.Due to the presence of surgical solution and the condition of the returned sample, the damage is most likely related to customer handling.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|