TEPHA, INC. - 3005670760 GALAFLEX; BREAST - ABSORBABLE MESH Back to Search Results
Model Number 100035-09
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Purulent Discharge (1812)
Date of Event 04/30/2018
Type of Reportable Event Serious Injury
Event or Problem Description
As reported, patient was implanted with a galaflex mesh on (b)(6) 2017.On (b)(6) 2018 patient was diagnosed with left breast implant failure to integrate and dimensional collapse.On (b)(6) 2018 30 cc yellow thick fluid was aspirated in the office.On (b)(6) 2018 10 cc yellow lipid fluid was aspirated.As reported the galaflex scaffold was collapsed and not attached to skin flap anteriorly that incision and drainage performed.As reported, the treating surgeon identified the condition of the left breast device failure to integrate and dimensional collapse as not caused by the use of the galaflex (p4hb) implant used to treat the patient.
 
Additional Manufacturer Narrative
As reported, post implant of galaflex, the patient experienced fluid discharge and left breast device failure to integrate.The surgeon did not identify any causal relationship between the failure to incorporate being due to the galaflex implant used to treat the patient.In an abundance of caution, events for which relatedness cannot be completely ruled out are processed for mdr reporting.Review of manufacturing records shows product was manufactured to specification.The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.Not returned - remains implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALAFLEX
Common Device Name
BREAST - ABSORBABLE MESH
Manufacturer (Section D)
TEPHA, INC. - 3005670760
99 hayden ave suite 360
lexington MA 02421
Manufacturer (Section G)
TEPHA, INC. - 3005670760
99 hayden ave suite 360
lexington MA 02421
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
8005566756
MDR Report Key15326342
Report Number1213643-2022-00603
Device Sequence Number6314611
Product Code OOD
UDI-Device Identifier00855920006041
UDI-Public(01)00855920006041
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K140533
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2020
Device Model Number100035-09
Device Catalogue NumberGP0608
Device Lot Number170308
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/15/2021
Initial Report FDA Received Date08/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/21/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
Patient SexFemale
-
-