MAUDE Adverse Event Report: EXACTECH, INC. LOGIC TIBIA PS MOD INSRT SZ 5 13MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. LOGIC TIBIA PS MOD INSRT SZ 5 13MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 02-012-35-5013
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Failure of Implant (1924); Insufficient Information (4580)
Date of Event 08/04/2022
Type of Reportable Event Serious Injury
Event or Problem Description
As reported, approximately 13 years post op the initial tka, this male patient was revised due to poly wear.Patient was revised to a logic psc size 5 15mm insert.
 
Additional Manufacturer Narrative
Pending evaluation.
 
Additional Manufacturer Narrative
Section h10: (h3) the revision reported in (b)(4) was likely the result of prosthesis wear of the tibial insert and patella after being implanted for approximately 13 years, possibly due to malalignment between the implants, high contact stresses during knee flexion, third body wear, patient-related conditions, instability, or any combination of these possibilities.The cause of the prosthesis wear cannot be determined because the revised components were not returned for evaluation and no radiographs were provided.An additional contributing factor to the revision may have been inclusion of the implanted tibial insert in the packaging recall.Section h11: the following sections have corrected information: (h6) component code: 734, bearings.
 
Event or Problem Description
According to the information provided, the patient underwent a left total knee replacement surgery.Approximately 12 years and 11 months later, the patient was revised.It appears that they are alleging prosthesis wear of an implanted polyethylene device.Components were not returned evaluation, no images, radiographs, and relevant clinical information were provided.No further patient impact was reported.No additional information was provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b1, b2, g2 (initial report), h3 (remove erroneously entered statement from -1 follow up report), h6: clinical code, types of investigation, investigation findings and investigation conclusion., h11 (correcting erroneous summary from -1 follow up report.H11: the reason for the revision reported in event cannot be confirmed from the information provided but may be the result of prosthesis wear or due to inclusion of the polyethylene in the packaging recall.Additionally, these devices appear to have been implanted for over ten years prior to the reported event.Additionally, a contributing factor to the reported wear may have been the result of being packaged in a non-conforming vacuum bag for more than five years.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and relevant clinical information were not provided if any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay corrected information.The following sections were corrected: h6: types of investigation, investigation findings and investigation conclusion, h10, h11.H10: multiple mdr reports were filed for this event, please see associated report: 1038671-2022-00976.The revision reported was likely the result of prosthesis wear of the tibial insert and patella after being implanted for approximately 13 years, possibly due to malalignment between the implants, high contact stresses during knee flexion, third body wear, patient-related conditions, instability, or any combination of these possibilities.The cause of the prosthesis wear cannot be determined because the revised components were not returned for evaluation and no radiographs were provided.An additional contributing factor to the revision may have been inclusion of the implanted tibial insert in the packaging recall.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC TIBIA PS MOD INSRT SZ 5 13MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 ct
gainesville FL 32653
Manufacturer Contact
kate jacobson
3523771140
MDR Report Key15326380
Report Number1038671-2022-00976
Device Sequence Number5374727
Product Code JWH
UDI-Device Identifier10885862001948
UDI-Public10885862001948
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K033883
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 03/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/04/2014
Device Catalogue Number02-012-35-5013
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/09/2022
Supplement Date Received by Manufacturer06/22/2023
12/12/2025
03/25/2026
Initial Report FDA Received Date08/31/2022
Supplement Report FDA Received Date07/12/2023
12/31/2025
03/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/06/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
-
-